From: Quality control material for the detection of somatic mutations in fixed clinical specimens by next-generation sequencing

Fig. 1

Limit of Detection (LoD) and linearity of the CHP2 assay. DNA isolated from four fresh-frozen cell line pellets was further diluted in DNA isolated from a de-identified normal human placenta. Variants were identified at the expected frequencies, down to 3 % mutant DNA in the context of normal DNA for fresh-frozen samples, when barcoding up to 7 samples. Pearson correlation is shown. Dotted line denotes the equality line